Calibration Solutions


Conductivity Standard Solutions and pH Buffers are used for factory calibration of water quality testers. Regular use of these solutions is recommended to ensure specified instrument accuracy. Frequency of conductivity recalibration depends upon use, but once every month should be sufficient for an instrument used daily. pH models, depending upon use, should be recalibrated with pH 7 Buffer every 1-2 weeks, and checked with pH 4 and/or 10 Buffers at similar intervals.
TDS Calibration Solutions by Myron L
ORP Standard Test Solutions by HannaORP standard solutions allow you to test the precision of ORP electrodes. For example, by immersing the electrode in HI 7020 solution, you should read within the 200 to 275 mV range (@20°C/68°F). If the reading is outside the indicated interval, clean and condition your ORP electrode in HANNA® pretreatment solution.
|
![]() |
Benefits of Calibration and Buffer Solutions
- Easy to Use
- Calibrating Water Treatment Instruments Regularly Ensures the Accurate Test Results
- Myron L solutions are NIST Traceable
- Myron L Company Standard Solutions have an accuracy of ±1% based on values published in the International Critical Tables
- pH Buffer Solutions 4, 7 and 10 are mold inhibited and accurate to within + 0.01 pH units @ 25°C
Calibration and Buffer Solutions Features
- Myron L solutions are NIST Traceable
Applications for Calibration and Buffer Solutions
|
|
|